Answer:
Tube MR = = 2.2851 x 10⁻³, Phase lag = 2.467°, Transducer MR = 1, Phase lag = 8.049°, b. System is acceptable, explained further
Step-by-step explanation:
a. Magnification Ratio = 1/[√{(1 – (W/W(n))²)² + (2εW/W(n))²}]
For tubing, W(n) = 30rad/s, ε = 0.45, W = 2π x 100 = 628 rad/s
M.R = 1/√[{1 – (628/30)²}² + (2 x 0.45 x 628/30)²]]
= 1/√(191147.7262 + 354.9456)
= 2.2851 x 10⁻³
Phase lag = tan-1 [(2εW/W(n))/1 – (W/W(n))2²]
= tan⁻¹ (18.84/437.204) = 2.467 °
For the transducer, W(n) = 6280 rad/s, W = 628 rad/s, ε= 0.70
M.R = 1/√[{1 – (628/6280)²}² + (2 x 0.7 x 628/6280)²}]
= 1/√[{1 – (628/6280)²)² + (2 x 0.7 x 628/6280)²]
= 1/√[(99/100)² + 0.142] = 1
Phase lag = tan⁻¹ [(2εW/W(n))/1 – (W/W(n))2]
Phase lag = tan⁻¹ [(2 x 0.7 x 628/6280)/1 – (628/6280)²]
Phase lag = tan⁻¹ (0.14/0.99) = 8.049°
The system is working far away from resonance point, so it can be used without any problem
b. (30 x 0.45) (tube) < (6280 x 0.7) (transducer)
This means that the transducer will come to a vibration- less state much quicker than the tube, hence, the pressure measuring system will be quick and accurate. It shows the good suitability of the taken system to the working system.