Answer:
1.022ppm is the unknown concentration of the metal
Step-by-step explanation:
Based on Lambert-Beer law, the increasing in signal of a detector is directly proportional to its concentration.
The unknown concentration (X) produces a signal of 0.255
99mL * X + 1mL * 100ppm / 100mL produces a signal of 0.502
0.99X + 1ppm produce 0.502, thus, X is:
0.255 * (0.99X + 1 / 0.502) =
X = 0.503X + 0.508
0.497X = 0.508
X =
1.022ppm is the unknown concentration of the metal