379,984 views
6 votes
6 votes
A pressure transducer has the following specifications: Input rage: 0-100 psi and the corresponding Output Range: 0-5 Volts. Linearity Error: 0.10% of the Reading Hysteresis Error: 0.10% of the Reading Sensitivity Error: 0.15% of the Reading Zero Drift Error: 0.20% of the Reading Its output is read via a voltmeter with instrument error of 0.10% of the reading and resolution of 0.01 V. If the applied pressure on the transducer is 65 psi, what is the design stage uncertainty of this pressure measurement system?

User Xiaoou Wang
by
2.6k points

2 Answers

15 votes
15 votes

Answer:

0.287

Step-by-step explanation:

User Gyandeep Sharma
by
2.3k points
20 votes
20 votes

Answer:

0.287

Step-by-step explanation:

Design-stage uncertainty can be expressed as :

Ud = √ Uo^2 + Uc^2 ------ ( 1 )

where : Uo = 1/2( resolution value ) = 1/2 * 0.01 V = 0.005 V

Uc = √(0.10)^2 + (0.10)^2 + (0.15)^2 + (0.20)^2 = 0.287

back to equation 1

Ud = √ ( 0.005)^2 + ( 0.287 )^2 = 0.287

User Andre Romano
by
2.6k points