Final answer:
The most likely issue after PM on a coagulation analyzer that results in PT control being outside 6 SD is the reset of the ISI value, which needs to be re-entered correctly for accurate test results.
Step-by-step explanation:
When troubleshooting the issue in a coagulation department where quality control (QC) for prothrombin time (PT) is outside of 6 standard deviations (SD), several factors need to be considered. Since the QC for activated partial thromboplastin time (aPTT) is within 2 SD, this suggests that the maintenance procedures may have specifically affected the PT test system. The problem could be attributed to changes in the calibration settings such as the International Sensitivity Index (ISI) value, which may have been reset or modified during the preventive maintenance. It is essential for the ISI value to be correctly entered to ensure accurate PT results. Option (b), which identifies the need to re-enter the ISI value, is the most probable cause for the discrepancy observed. It is not correct to assume that maintenance will invariably alter the PT range (option d), and it is a requirement to perform QC after maintenance (opposing option a), so skipping QC is not advisable. Similarly, assuming the controls are contaminated without further evidence is premature (option c).