23.2k views
4 votes
What can be done when investigating a possible kidney stone to better

demonstrate shadowing from the stone when using a 3 MHz probe?
a. Position the focal point deeper than the stone.
b. Increase the amount of gain.
c. Increase the frequency to 5 MHz.
d. Decrease the output power of the system.

1 Answer

4 votes

Final answer:

Enhancing shadowing of a kidney stone in ultrasound is best achieved by positioning the focal point deeper than the stone when using a 3 MHz probe since higher frequencies do not penetrate as well into the tissue.

Step-by-step explanation:

When investigating a possible kidney stone using ultrasound with a 3 MHz probe, the intention is to create an image with sufficient detail to identify and evaluate the stone. It's important to understand that the frequency of ultrasound has a direct impact on both the resolution of the image and the depth of penetration. Higher frequencies provide greater detail but have limited penetration into tissue. Conversely, lower frequencies penetrate deeper but offer less detail.

Given the depth limitation of the tissue, if the kidney stone is located deeply, it may be beneficial to position the focal point deeper than the stone; however, increasing the frequency to 5 MHz as stated in option c might not be the best approach due to reduced penetration depth. Therefore, option a, positioning the focal point deeper, is the most appropriate method to enhance the shadowing of the kidney stone as it optimizes the ultrasound beam for the depth of the object of interest. Adjusting the gain and decreasing the output power do not directly impact image resolution in relation to shadowing from the kidney stone.

User Wolfhammer
by
8.2k points