Which of the following grids used in x-ray production will require the highest mA-s setting?

A. 2 to 1
B. 4 to 1
C. 8 to 1
D. 12 to 1

User Mbonness
by
7.8k points

1 Answer

4 votes

Final answer:

The 12 to 1 grid ratio in x-ray production will require the highest mA-s setting due to its increased ability to absorb scattered radiation and maintain image quality.

Step-by-step explanation:

The question is asking which of the following grids used in x-ray production will require the highest milliampere-seconds (mA-s) setting: A. 2 to 1 B. 4 to 1 C. 8 to 1 D. 12 to 1.

The answer to this question is D. 12 to 1. Grids with higher grid ratios, like the 12 to 1 grid, tend to absorb more scattered radiation and therefore require an increase in the mA-s to maintain image quality.

The grid ratio describes the ratio of the height of the lead strips to the distance between them, with a higher ratio indicating taller strips and less space for scatter to get through.

As the grid ratio increases, the ability of the grid to clean up scatter improves, but the more radiation it absorbs, necessitating an increase in x-ray exposure to compensate, hence a higher mA-s setting is needed to produce a diagnostic-quality image.